Universitas Dian Nuswantoro (UDINUS), through researchers Kurniatie, Menik Dwi; Asmaria, Talitha; Rokhmanto, Fendy; Triwardono, Joko; Senopati, Galih; Mayasari, Dita Ayu; and Anandisa, Wulan Tsabita, conducted a study to reproduce a commercial aneurysm clip using metal forming. The activity concerned an aneurysm clip used as the primary treatment for complex brain aneurysms, particularly in haemorrhagic stroke associated with blockages or ruptures of cerebral blood vessels. These conditions can lead to severe neurological deficits.
The stated aim was to reproduce the commercial aneurysm clip through metal forming so that medical equipment could be provided more independently throughout the country. The source also identifies open surgical clipping in some developing countries, including Indonesia, as sometimes unaffordable and a financial burden to the state.
The metal-forming process employed rolling and bending techniques. A base and a jig system were used to press Ti6Al4V wire into an R-shaped contour. The investigation examined structural integrity, fine-grain structure, mechanical properties, density, chemical composition, and closing force. The source reports that the structural integrity, fine-grain structure, and mechanical properties were confirmed. After forming, the aneurysm clip had a density of 4.3 g/cm3. Chemical composition testing used a scanning electron microscope equipped with an energy-dispersive X-ray spectrometer and revealed Ti, Al, and V elements.
The fabricated clip exhibited a closing force of 1.689 N, compared with 0.791 N for the commercial aneurysm clip. The source describes the fabricated clip's closing force as about double that of the commercial clip. This UDINUS activity demonstrates health-related research and metal-forming research directed toward reproducing a medical device.